ISTARI
Model 1

Enclosure for extended care and critical care interventions in a hospital environment for patients who require airborne, droplet, or contact isolation precautions.
Design Features
- Footprint sized for use with a typical hospital bed and commode in a U.S. hospital room
- Clear plastic walls and replaceable modular components compatible with decontamination and disposal methods
- Integrated air intake filters and HEPA outflow filter attached to blower fan
- Patient access via the HugSuit and lean-in windows at head and side of bed
- Equipment access via conduit panel for wires and tubing
- Airlock pass-throughs
Technical Specs
- Design Specification Requirements / Operational Airflow = Device shall perform at least 12 air exchanges per hour
- Performance is ~20 air exchanges/hour
- HEPA (99.97% particulates @ 0.3μ)
- Access via conduit panel for wires and tubing
- Airlock pass-throughs
Coming Soon!
Coming Soon!
Model 1 In Action
Clinical simulation activities have been integral to the iterative design evolution of the ISTARI Model 1. Utilizing advanced simulation facilities at the UNMC Global Center for Health Security and engagement with Innovation Test Bed partner hospitals, diverse healthcare teams have provided valuable feedback to refine device features, optimize clinical and IPC utility, and improve patient safety and experience. A broad scope of medical interventions has been evaluated, including emergency and critical care procedures.
ISTARI Model 1
3D Model
WATCH
ISTARI IN ACTION
ISTARI is designed to allow for the full spectrum of clinical care and is compatible with all transmission-based precautions. The units are compatible with standard medical utilities in a hospital room. It can be used in existing hospital rooms or in temporary surge spaces with portable utilities. See it in action…